Code: LC012
Features
Features (All Models)
-100% exhausted design for work involving high volumes of volatile chemicals
-Long-life DC Electronically Commutated Motor (ECM)
-BalanceFlowTM plenum provides energy-efficiency and ultra-quiet operation.*
-Intrinsically-safe negative pressure design
-Mag-Zone™ area on interior right side manages storage challenges with Labconco Magnetic Mag-Zone accessories
-Constant Airflow Profile™ (CAP) airflow monitoring system
-Air-Wave™ Entry System at access opening to enhance containment
-Contain-Air™ Negative Pressure Channel at top of sash to enhance containment
-Includes airflow sensor to display inflow and downflow on display
-99.99+% efficient supply and exhaust HEPA filters. 99.999% efficient ULPA filters or H14 HEPA filters also available
-Logic OS and 5″ interior-mounted, line-of-sight touchscreen information center
-Smart-Start™ System allows user to start up and shut down cabinet functions by raising and lowering the sash
-Full operational maintenance and certification logs for detailed record keeping
-Status line for alarm conditions and alerts to warn when filter life diminishes to 20% and 10% before alarming at 0%
-Interval/elapsed timer for experiment monitoring
-UV light timer (on models with UV)
-Password protection for cabinet settings
-8 Languages
-Remote Communication Board for direct communication with remote blower and/or building management system
-Dry contact relays
-USB data output
-RS-232 RJ45 data output
-Ethernet pass-through
-Bright, glare-free LED lighting up to 180 foot-candle (1937 lux). Located outside the contaminated work area
-10° angled, fully closing sash. 1/4″ tempered safety glass sash with counterbalanced, anti-racking, easy-to-lift mechanism
-Curved, stainless steel inlet grille with waterfall design and Reserve-Air™ Secondary Airflow Slots
-Removable, seamless type 304 stainless steel, dished work surface with lift out knobs
-Two electrical outlets. Flush mounted stainless steel cover with dampened hinges.
-115V GFCI, double receptacles, maximum amperage draw 5A.
-230V, single receptacles, maximum amperage draw 3A.
-Unitized, pressure tested stainless steel interior. Optic White epoxy-coated steel exterior
-Removable towel catch located under work surface
-8″ (20.3 cm) working sash height
-22.5″ (57 cm) max. sash opening & 27.0″ (69 cm) viewing height
-Airtight damper with 10″ (25.4 cm) nominal round connection
-Noise level <67 dBA
-Nominal inflow velocity of 105 fpm (0.53 m/sec)
-Nominal downflow velocity of 55 fpm (0.28 m/sec)
-10′ (3 m) power cord with plug
-Six year warranty
Options
4’ and 6’ nominal widths
International electrical configurations
Unassembled, NSF-approved, epoxy-coated steel, non-welded telescoping base stand with leveling feet
Accessory Package
UV lamp (254 nm)
Right-side service fixture with ball-type valve
Right and left-side Vacu-Pass™ Cord & Cable Portal
Cell Logic Packages
Scope-Ready™ Package consisting of 7.0″ wide x 10.5″ high (18 x 27 cm) sash cutout and Pure-Vu™ Seal with powder-coated steel collar. Stand-Still™ Isolation Platform is required and sold separately.
Temp-Zone™ Package consisting of temperature-controlled work area, 10.5″ wide x 12.0″ deep (27 x 30 cm) located on the left side of the work surface. Water circulator is required (not included). Circulator requirements: minimum flow rate of 0.4-0.5 GPM (1.5-1.9 LPM), typical volume of 6 liters, and temperature of 0-50° C (chiller/heater).
Required Accessories (Not Included)
Dedicated ductwork and remote blower required for each Type B2 BSC (CETA CAG-007)
Supporting base if non-welded stand option is not selected
Stand-Still™ Isolation Platform for Scope-Ready™ Cell Logic Models
Compliance
NSF/ANSI Standard 49 listed
NSF/ANSI 49 compliant (Cell Logic models)
ISO Class 5 (Grade A) per ISO 14644-1
UL Standard 61010-1
CAN/CSA 22.2 No. 61010-1
CE/UKCA Mark (230V models)
ADA-compliant
Reviews